Temperature-sensitive products moving through regional distribution may pass through several transport and handling stages before they reach a receiving or quality team. MaxiLog PDF is a pre-programmed, single-use temperature data logger that records shipment conditions so that temperature history can be reviewed after delivery.

It measures from -30°C to +70°C and records temperature every 10 minutes for up to 90 days, with capacity for 14,400 data points. Its compact 94 mm × 20 mm × 7 mm format is suitable for small packages, while the standard 30-minute start delay allows the logger to adjust to the cargo’s internal temperature before recording begins.

At the end of the trip, connect the built-in USB-A interface to a computer. The logger provides PDF and CSV reports, with the PDF presenting a graph and summary of the recorded data. Receiving or quality teams can then review the temperature record against the shipment’s required temperature conditions.

  • Single-use, pre-programmed monitoring for one shipment with a standard 30-minute start delay.
  • Automatic report access through the built-in USB-A interface, with PDF and CSV files and no dedicated logger software required for retrieval.
  • Up to 90 days of recording at 10-minute intervals, with storage for 14,400 data points
  • Compact 94 mm × 20 mm × 7 mm format for use inside small packages.
  • Programmable high and low alarm limits, with green operating status and a red flashing alarm indication.
  • Push-button Start/Mark/Stop operation, with up to 10 event markers shown on the report graph.

MaxiLog PDF is suitable for temperature monitoring in shipment applications such as:

  • Pharmaceutical products
  • Vaccines and biologics
  • Life-science shipments
  • Refrigerated food products
  • Temperature-sensitive chemicals
  • Regional cold-chain distribution and post-delivery shipment review
  • Model: 7050-G2 MaxiLog PDF V3 Temperature Data Logger
  • Recording Type: Single Use
  • Temperature Measurement Range: -30°C to +70°C
  • Temperature Accuracy: ±0.3°C
  • Temperature Resolution: 0.1°C
  • Data Storage Capacity: 14K data points (14,400)
  • Battery Life / Type: CR2032 lithium coin cell, 18-month shelf life
  • Data Sampling Interval: 10 minutes
  • Start-Up Delay: 30 minutes
  • Recording Duration: 90 days
  • Alarm Programming: Programmable high and low limits; alarm triggered when limits are exceeded
  • Alarm Modes: Exceed Limit
  • Alarm Reset Options: Single Use: No reset option
  • Alarm Notifications: Red LED flashes when alarm has been triggered
  • Indicators: Green LED for on/recording; red LED for alarm
  • Start-Up Options: Push button: Start/Mark/Stop
  • Dimensions: 94 mm (H) × 20 mm (W) × 7 mm (D)
  • Weight: 10.6 g / 0.37 oz.
  • Quality Assurance Certifications: CE, UKCA, RTCA, NIST Traceable Certificate of Calibration, IP67 and RoHS compliant
  • Minimum System Requirements: Windows XP, Vista, 7, 8 and 10; Mac and Android
  • Report Formats: PDF and CSV
  • Required Software: PDF Reader
  • Computer Interface: USB A, built-in interface
  • Warranty: 1 year
What does MaxiLog PDF record?

It records shipment temperatures from -30°C to +70°C at 10-minute intervals for up to 90 days.

How are the reports accessed after delivery?

Connect the built-in USB-A interface to a computer. The logger provides the recorded data in PDF and CSV formats.

Does MaxiLog PDF provide live or remote temperature monitoring?

No. The temperature record is reviewed after the logger is physically connected to a computer by USB.

Can MaxiLog PDF be reused?

No. MaxiLog PDF is a single-use temperature data logger.

Can MaxiLog PDF automatically decide whether a shipment should be released or accepted?

No. It provides a temperature record and alarm information for review. Release or acceptance decisions should follow the user’s own product requirements and procedures.

Does the logger require dedicated software?

No dedicated logger software is required to retrieve the report. A PDF Reader is required to view the PDF file.

Important: Do not open the sleeve bag until the trip is finished and you are ready to download the report.

1. Start the logger: Press and hold the button for at least 3 seconds until the OK LED flashes 5 times. The OK LED then flashes every 2 seconds to indicate that the logger is operating. The standard 30-minute start delay runs before temperature recording begins.

2. Set a mark if needed: Double-click the button. The OK LED flashes 3 times to confirm the mark. Up to 10 event marks can be added, and each mark appears on the graph at the corresponding point in time.

3. Stop the logger: Press and hold the button for at least 3 seconds until the Alarm LED flashes 5 times. Alternatively, connect the USB interface to a computer to stop the logger automatically.

4. Get the report: Connect the USB interface to a computer. You may need to wait until both LEDs become solid. The logger opens as a flash drive containing the data report in PDF and CSV formats.

TempU Temperature Recorder Product Brochure: Download